Neutral Gas Ion Mass Spectrometer (NGIMS) Pre-Kiwi Test FC Analog and FC CPU Test and Verification Procedure For CONTOUR NGIMS Test Hardware Test Prepared by: _________________________ Florence Tan NGIMS Digital Engineer Approved by: _________________________ Jack Richards NGIMS Instrument Engineer 1.0 Objective This document describes the procedure for: * Testing and verifying the NGIMS FC and Analog electronics 2.0 References 2.1 S. Williams CONTOUR 1553 Bus Specifications Version 0.92 2.2 R. Miller NGIMS Spacecraft Simulator Use's Manual 077-0432 2.3 M. Paulkovich Software Requirememts Document, GSFC/NGIMS-FSW-011, CONTOUR/NGIMS Telemetry, GSFC/NGIMS-FSW-06, Functional Requirements for CONTOUR/NGIMS CDS Low-Level Interface Flight Software GSFC/NGIMS-FSW-12 2.4 Test Setup NGIMS FC CPU breadboard NGIMS FC Analog Breadboard NGIMS Spacecraft Simulator HP1660CS logic analyzer NGIMS FC debug board ??? ADA tools for software development [to be added] NGIMS GSE workstation 3.0 Overview 3.1 Software Overview 3.2 Spacecraft Simulator Overview 4.0 Test Procedure 1. FC Analog and FC CPU interface test. a) Check pin assignments - make sure power thru FC Analog b) Turn on Power c) Record current readings d) Record +15/-15/+5 voltages (need to understand V drops) e) Verify no data flow f) Set encounter bit to enable data g) Record outputs for following AMUX channels: Channel 81 - Imon 2 Channel 82 - HV1 Mon Channel 83 - HV5 Mon Channel 87 - B- Mon Channel 88 - FC Temp Channel 89 - +5 Mon Channel 90 - +15 Mon Channel 91 -15 Mon h) Send commands on DCON and record outputs on: DCON1-4 OS Fil1 On and verify on AMUX input 1 DCON1-5 OS Fil2 On and verify on AMUX input 2 DCON1-6 CS Fil1 On and verify on AMUX input 3 DCON1-7 CS Fil2 On and verify on AMUX input 4 DCON1-8 BA Fil On and verify on AMUX input 5 Other data from FC AMUX includes: i) Send command from Sim to FC CPU to set DCON and see the values on Logic Analyzer: SIGNAL Use Module DCON1-0 OS-LOW-EMIS\ OS DCON1-1 OS_ADJB\ OS DCON1-2 CS_LOW_EMIS\ CS DCON1-3 CS_ADJB\ CS DCON1-4 OS_FIL1_ON OS DCON1-5 OS_FIL2_ON OS DCON1-6 CS_FIL1_ON CS DCON1-7 CS_FIL2_ON CS DCON1-8 BA_ON BA DCON1-9 BA_LO EMIS BA DCON1-10 EM1_SEL EM DCON1-11 EM2_SEL EM DCON1-12 DCON-SP4 BA DCON1-13 DCON-SP1 MB DCON1-14 DCON-SP2 MB DCON1-15 DCON-SP3 MB SIGNAL Use Module DCON2-0 Valve 1+ (ON only) BA DCON2-1 Valve 1- (ON only) BA DCON2-2 Valve 2+( ON only) BA DCON2-3 Valve 2- (ON only) BA DCON2-4 SPARE_TO_RB1 RB DCON2-5 SPARE_TO_BA1-1 BA DCON2-6 SPARE_TO_OS-1 OS DCON2-7 SPARE_TO_OS-2 OS DCON2-8 CS-LOW-EMIS\ CS DCON2-9 SPARE_TO_CS-2 CS DCON2-10 DCON_SP-4 MB DCON2-11 NC DCON2-12 NC DCON2-13 NC DCON2-14 NC DCON2-15 NC 2. FC with released software verified at GSFC a) Send commands from SIM to FC CPU to set DACs (flat table) b) Record voltage settings on FC Analog 3. Check MB pin assignments. 915-NGIMS-KIWI-001.V0 NGIMS pre-KIWI Test Document Rev.: 0.0 Date: Oct 9 2000 Page: 1