;ngims_postlaunch_PressureChkHi.txt ;ft Rev 1 July 8, 2002 ;power up NGIMS using NGIMS_PwrOn.prc ;NGIMS_TM_ON $tm on ;NGIMS_LOADMT_B3 $LoadMT 18 ;AMUX table with fil mons at 5x rate $WAIT 4 ;The higher rate gives better time resolution for turnon. ;NGIMS_Pressure_Test_Part0 $tableupdate ;Load RAM with EEPROM values incl. hi file sampling $Wait 5 ;Table update turns off all DAC Overrides $FLast 2580 ;0 volts DACs $FFirst 2566 ;All DAC output voltages at 0 (except OS EFB1/2) $Wait 2 $dac 8 0 $dac 9 0 $DACOR 8 ON ;DACOR for EM1 On (DAC 8 = 0 during Elec short test) $DACOR 9 ON ;DACOR for EM2 On (DAC 9 = 0 during Elec short test) $Hidestrip m81 $strip m81 e .5 1.5 0 0 500 200 ;Imon Normal - was set to .6A FS $strip m65 e 0 2.5 902 233 200 200 ;BA Fil Amps $strip m66 e 0 1000 902 433 200 200 ;BA Grid Emission $strip m67 e -100 500 902 633 200 200 ;BA Out, 500 pico amps fs $Wait 5 ;Wait for strip chart display ; ;*******************BA FILAMENT ON ****************** ;NGIMS_BAFil_On $DCON0 8 1 ;BA filament on ;NGIMS_Pressure_Test_Part2 $NOP 1 370 ;Start plot for BA monitors and end for Imon short plot. $DCON0 9 1 ;BA set Hi Emission $WAIT 59 ;Wait for filament turnon before limitcheck ;Operator enter (but do not send) INITMODE in cmnd window. $NOP 1 375 ;NGIMS_Pressure_Test_Part3 $NOP 1 380 $WAIT 5 $DCON0 8 0 ;BA filament OFF $DCON0 9 0 ;Set BA back to Lo Emission $WAIT 10 $NOP 1 385 ;End BA plots $WAIT 5 ;Wait for NOP ;power off NGIMS using NGIMS_PwrDn.prc