; This script continues the BA pressure check in low emission ; by going to high emission. BA filament is on but turned on again for standaloneness. ; FT 5/9/02 Initial rev $TM ON $tcrate 1000 ;Set comnd rate at 1/sec. Comment out for STOL convert $NOP 1 370 $WAIT 5 ;Wait for NOP $DCON0 8 1 ;BA filament on $DCON0 9 1 ;BA Hi Emission $WAIT 30 ;Wait for BA hi before limitcheck $NOP 1 375 $Bell $Bell $Bell $HOLD ;If BA Fil MORE THAN 2.4A send Initmode - immediately ;BA Fil __________(2.0A) BA Grid _________(527 uA) ;BA Out _________(<1000 pA) 50 pA on 2-Dec-01 ;If BA-Out is less than 1000 pA Operator send GO ;100 pA ~= 1e-7 mbar (Hi emission) ;Operator clear INITMODE from command window.